Remote Control Commands
R&S
®
SMW200A
1014User Manual 1175.6632.02 ─ 16
Table 13-10: Value range
Impairments Min [dB] Max [dB] Resolution
Digital -30 30 0.01
Analog -10 10 0.01
Parameters:
<Angle> float
Range: -30 to 30
Increment: 0.01
*RST: 0
Default unit: DEG
Example:
SOURce:BB:IMPairment:RF1:QUADrature:ANGLe -5
Sets the quadrature offset to -5 degrees.
Manual operation: See "Quadrature Offset" on page 412
[:SOURce]:BB:IMPairment:BBMM<ch>:SKEW <Skew>
[:SOURce]:BB:IMPairment:FADer<ch>:SKEW <Skew>
[:SOURce]:BB:IMPairment:IQOutput<ch>:SKEW <Skew>
[:SOURce]:BB:IMPairment:RF<ch>:SKEW <Skew>
Sets a delay between the Q vector and the I vector of the corresponding stream.
Parameters:
<Skew> float
Range: -500E-9 to 500E-9
Increment: 1E-12
*RST: 0
Manual operation: See "Skew" on page 412
[:SOURce<hw>]:IQ:IMPairment[:STATe] <State>
[:SOURce]:BB:IMPairment:BBMM<ch>:STATe <State>
[:SOURce]:BB:IMPairment:FADer<ch>:STATe <State>
[:SOURce]:BB:IMPairment:IQOutput<ch>:STATe <State>
[:SOURce]:BB:IMPairment:RF<ch>:STATe <State>
Activates the impairment or correction values LEAKage, QUADrature and IQRatio
for the corresponding stream.
Parameters:
<State> 0 | 1 | OFF | ON
*RST: 0
Example:
SOURce1:BB:IMPairment:RF1:STATe ON
Activates digital impairment.
Manual operation: See "State" on page 411
SOURce Subsystem