Remote Control Commands
R&S
®
SMW200A
1137User Manual 1175.6632.02 ─ 16
:SYSTem:REBoot
Reboots the instrument including the operating system.
Usage: Event
:SYSTem:RESTart
Restarts the instrument without restarting the operating system.
Usage: Event
:SYSTem:SHUTdown
Shuts down the instrument.
Usage:
Event
:SYSTem:WAIT <TimeMs>
Delays the execution of the subsequent remote command by the specified time.
This function is useful, for example to execute an SCPI sequence automatically but
with a defined time delay between some commands.
See Chapter 11.3, "Assigning Actions to the User Key", on page 678.
Setting parameters:
<TimeMs> integer
Wait time in ms
Range: 0 to 10000
*RST: 0
Example:
:SYSTem:WAIT 10000
// waits 10s before resetting the instrument
*RST
Usage: Setting only
Manual operation: See "Wizard" on page 680
13.20 STATus Subsystem
This system contains the commands for the status reporting system. See also Chap-
ter B.1.5, "Status Reporting System", on page 1203 for detailed information.
*RST on page 772 has no effect on the status registers.