Monitoring Signal Characteristics
R&S
®
SMW200A
594User Manual 1175.6632.02 ─ 16
Example: Results calculation at the RF output
The reference value for the power calculation at the RF outputs is the "Level" value, as
displayed in the status bar. A configured power offsets, for example "RF > RF Level >
Offset" are considered.
Set the "RF Level = 5 dBm"
In the "Setup > Baseband Powers Simulation" dialog, enable a power measurement
with the following properties:
●
"Source > Baseband A"
●
"Output > RF A"
●
"Acquisition > Nominal"
The display shows "PEP = 15 dBm", "Level = 5 dBm"
8.2.3 Baseband Powers Settings
Access:
► Select "Taskbar > System Config > Setup > General > Baseband Powers".
The dialog provides a summarized information on the important baseband level
parameters, like power level and peak power (PEP) of selectable acquisition
points.
The remote commands required to define these settings are described in Chap-
ter 13.18.4.9, "SOURce:BB:MEASurement Subsystem", on page 1019.
Querying Information on the Baseband Power Levels