Configuring the Signal in the RF Domain
R&S
®
SMW200A
526User Manual 1175.6632.02 ─ 16
Risk of DUT damage due to high input power
The settings combination "ALC > Sample Level > Full" and "ALC > State > Off (Sample
&Hold)" or "ALC > State > Sample & On" results in signals with large headroom. With
this settings combination, the R&S SMW generates a short sample pulse with full RF
output power.
To protect the DUT from damage due to high input power, use the setting "Sample
Level > Attenuated" or "Sample Level > Minimum".
7.8.2.1 Required Options
The equipment layout for automatic level control includes per signal path
Option frequency (R&S SMW-B10x)
7.8.2.2 ALC Settings
Access:
Select "RF" > "RF Level" > "Automatic Level Control".
In the "ALC" dialog, you can configure the settings for the automatic level control of
the RF signal to achieve optimal accuracy.
The remote commands required to define these settings are described in Chap-
ter 13.18.15, "SOURce:POWer Subsystem", on page 1092.
Improving Level Performance