See also Chapter 8, "Monitoring Signal Characteristics", on page 559.
2.3.7 Saving and Recalling Settings
To restore the results of our measurements later, we store the instrument settings to a
file.
To save the instrument settings to a file
We assume, a test configuration as described in Chapter 2.3.5, "Routing the Signal
through the Instrument and Defining the Output Connectors", on page 78.
1. Press the SAVE/RCL key on the front panel.
2. In the "Save/Recall" dialog box, select "Operation Mode > Save".
Tap the "Filename", use the on-screen keyboard, and enter MyTestSignal.
Trying Out the Instrument