Configuring the Internal Baseband Source
R&S
®
SMW200A
314User Manual 1175.6632.02 ─ 16
Remote command:
[:SOURce<hw>]:BB:ARBitrary:TSIGnal:RECTangle:AMPLitude on page 929
Offset DC
Enters a DC component.
Remote command:
[:SOURce<hw>]:BB:ARBitrary:TSIGnal:RECTangle:OFFSet on page 929
Generate Signal HD
Generates a signal and saves it as a waveform file on the hard disk. Use the standard
"File Select" function to store this file.
Remote command:
[:SOURce<hw>]:BB:ARBitrary:TSIGnal:RECTangle:CREate:NAMed
on page 931
Generate Signal RAM
Generates a signal and uses it as output straight away. The instrument stores the file
with a predefined name.
Remote command:
[:SOURce<hw>]:BB:ARBitrary:TSIGnal:SINE:CREate on page 931
4.7.3.4 Constant IQ Test Signal
Access:
1. Select "Baseband > ARB > General > Create Test Signal > Const IQ".
2. Select "Create Test Signal".
This dialog provides settings for configuration of a continuous test signal with con-
stant IQ. The I and Q values for each path are adjustable but constant. For more
information, refer to Chapter 4.7.2.2, "ARB Test Signals", on page 301.
Converted to analog, the signal is output directly at the I and Q connectors of the
instrument.
Using the Arbitrary Waveform Generator (ARB)