EasyManua.ls Logo

Rohde & Schwarz SMW200A - Establishing a Remote Control Connection over LAN Using VXI-11 Protocol

Rohde & Schwarz SMW200A
1312 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Network Operation and Remote Control
R&S
®
SMW200A
733User Manual 1175.6632.02 ─ 16
Starting a remote control via the LXI web browser
This section assumes that the instrument and the controller PC are connected in the
LAN.
1. Start a web browser that supports html5 (W3C compliant).
2. Enter the IP address of the R&S SMW in the browser's address bar.
The R&S SMW's welcome page is displayed.
3. In the navigation pane, select "Instrument Control" > "Web Control".
Remote access to the instrument requires the password. The default password is
instrument.
4. Enter the password and confirm with the ENTER key.
After the connection is established, the current screen of the R&S SMW is dis-
played in the browser window.
5. Use the mouse cursor and keyboard to access the functionality of the instrument
as you would directly perform on the instruments touchscreen and front panel.
12.8.2 Establishing a Remote Control Connection over LAN Using VXI-11
Protocol
In this example, the I/O software library R&S VISA from Rohde & Schwarz is used to
set up a LAN remote control link and remotely control the R&S SMW. R&S VISA is run-
How to Set Up a Remote Control Connection