Configuring the Internal Baseband Source
R&S
®
SMW200A
364User Manual 1175.6632.02 ─ 16
File ← Output Settings
Accesses the standard "File Select" function to specify the output filename of the multi-
carrier waveform to be calculated. As with normal waveforms, the file extension is
*.wv.
To trigger the calculation and storage of this multi-carrier waveform, select "Create" or
"Create and Load".
Remote command:
[:SOURce<hw>]:BB:ARBitrary:MCARrier:OFILe on page 947
Clock Rate ← Output Settings
Displays the resulting sample rate at which the multi-carrier waveform is output by the
ARB generator.
The output clock rate depends on the following:
Number of carriers
Carrier spacing
Input sample rate of the leftmost or rightmost carriers.
Remote command:
[:SOURce<hw>]:BB:ARBitrary:MCARrier:CLOCk? on page 948
File Size ← Output Settings
Displays the resulting number of samples of the multi-carrier waveform.
Remote command:
[:SOURce<hw>]:BB:ARBitrary:MCARrier:SAMPles? on page 945
Create/Create and Load
Creates a multi-carrier waveform defined by the multi-carrier table and general setting
entries.
If you start calculation with the "Create and Load" function, the instrument creates the
waveform and loads it in the ARB generator.
This multi-carrier waveform is stored with filename specified in "Output File". Depend-
ing on the configuration of the multi-carrier waveform, calculation can take some time.
To stop the calculation, use the "Abort" function.
Remote command:
[:SOURce<hw>]:BB:ARBitrary:MCARrier:CREate on page 948
[:SOURce<hw>]:BB:ARBitrary:MCARrier:CLOad on page 948
Generating Multi-Carrier Signals