Configuring the Signal in the RF Domain
R&S
®
SMW200A
522User Manual 1175.6632.02 ─ 16
The configuration "RF Off Mode > Unchanged" ensures that constant VSWR is
maintained if you turn the signal off and on again.
Highest level accuracy
Enter the target level value in "Fixed" mode and activate ALC (automatic level con-
trol)
In this operating mode, the generator adjusts the output level whenever the fre-
quency or level setting changes.
Noise sensitive applications
The parameter "RF OFF Mode > Full Attenuation" sets maximum attenuation and
thus suppresses noise when you turn off the RF signal.
Interactions and characteristics
The spectral purity of the RF output signal decreases with high attenuation.
Fixed mode is effective when automatic level control is activated (ALC State = On).
The setting of the RF OFF Mode > Full Attenuation has priority over the RF level
Mode > Fixed or Manual, regardless of the selected settings characteristics.
7.8.1.1 Attenuator Settings
Access:
Select "RF" > "RF Level" > "Level > Attenuator".
In the "Attenuator" dialog, you can select the operating mode of the electronic
attenuator, the instrument is equipped with.
The remote commands required to define these settings are described in Chap-
ter 13.15, "OUTPut Subsystem", on page 808 and Chapter 13.18.15,
"SOURce:POWer Subsystem", on page 1092.
Improving Level Performance