Remote Control Commands
R&S
®
SMW200A
812User Manual 1175.6632.02 ─ 16
Return values:
<Tripped> 0 | 1 | OFF | ON
*RST: 0
Example:
OUTP:PROT:TRIP
queries the state of the protective circuit of the RF output.
Response: 0
the protective circuit has not tripped.
Usage: Query only
Manual operation: See "Overload" on page 524
13.16 SENSe, READ, INITiate and SLISt Subsystems
These subsystems contain the commands for configuring the power measurements
with R&S NRP power sensor connected to the R&S SMW.
The local state is set with the INIT command. Switching off the local state enhances
the measurement performance. Measurements results can be retrieved in local state
on or off.
Sensor parameters are set with the SENSe commands.
To start the measurement and retrieve the result, use the :READ<ch>[:POWer]?
command.
Suffix Value range Description
SENSe<ch>
[1] to 4 Indicates the sensor
Sensor mapping:
●
SENSe[1] - default mapping for sensors connected to the
SENSOR connector
●
SENSe2 - sensor connected to a USB connector
●
SENSe3|4 - further connected sensors to USB connectors,
in the connection order
Use the :SLISt commands to change the sensor mapping
READ<ch>
[1] to 4 Sensor assignment
INIate<hw>
[1] to 4 Sensor assignment
ELEMent<ch>
[1] to 25 Sensor mapping list
SENSe, READ, INITiate and SLISt Subsystems