Remote Control Commands
R&S
®
SMW200A
910User Manual 1175.6632.02 ─ 16
[:SOURce<hw>]:BB:DM:TRIGger[:EXTernal]:INHibit <Inhibit>
Specifies the number of symbols by which a restart is inhibited.
Parameters:
<Inhibit> integer
Range: 0 to 21.47*symbRate
*RST: 0
Default unit: symbol
Example: See Example "Adjusting clock, marker and trigger settings"
on page 892
Manual operation: See "External / Trigger Inhibit" on page 257
Marker Settings
[:SOURce<hw>]:BB:DM:TRIGger:OUTPut<ch>:MODE <Mode>
Defines the signal for the selected marker output.
For detailed description of the regular marker modes, refer to "Marker Modes"
on page 235.
Parameters:
<Mode> CLISt | PULSe | PATTern | RATio
CLISt
A marker signal that is defined in the selected control list is gen-
erated.
*RST: RATio
Example: See Example "Adjusting clock, marker and trigger settings"
on page 892
Manual operation: See "Marker Mode" on page 259
[:SOURce<hw>]:BB:DM:TRIGger:OUTPut<ch>:ONTime <OnTime>
[:SOURce<hw>]:BB:DM:TRIGger:OUTPut<ch>:OFFTime <OffTime>
Sets the number of symbols in a period (ON time + OFF time) for marker RATio.
*)
If R&S SMW-B9 is installed, the minimum marker duration depends on the sample/
symbol rate.
See "Marker Minimum Duration" on page 236.
Parameters:
<OffTime> integer
Range: 1 (R&S SMW-B10) / 16* (R&S SMW-B9) to
16777215
*RST: 1
Default unit: symbol
SOURce Subsystem