Remote Control Commands
R&S
®
SMW200A
1106User Manual 1175.6632.02 ─ 16
[:SOURce]:ROSCillator[:INTernal]:ADJust[:STATe] <State>
Determines whether the calibrated (off) or a user-defined (on) adjustment value is
used for fine adjustment of the frequency.
Parameters:
<State> 0 | 1 | OFF | ON
0
Fine adjustment with the calibrated frequency value
1
User-defined adjustment value
The instrument is no longer in the calibrated state.
The calibration value is, however, not changed. The instrument
resumes the calibrated state if you send
SOURce:ROSCillator:INTernal:ADJust:STATe 0.
*RST: n.a. (factory preset: 0)
Example:
// query calibration value
CALibration:ROSCillator?
// 32767
// Set an internal source
// Activate user-defined adjustment value of 1000
SOURce:ROSCillator:SOURce INT
SOURce:ROSCillator:INTernal:ADJust:STATe 1
SOURce:ROSCillator:INTernal:ADJust:VALue 1000
// to resume calibrated state
SOURce:ROSCillator:INTernal:ADJust:VALue 0
SOURce:ROSCillator:INTernal:ADJust:STATe 0
// or
// SYSTem:FPRest
Manual operation:
See "Adjustment Active" on page 448
13.18.17 SOURce:SWEep Subsystem
The SOURce:SWEep subsystem contains the commands for configuring RF sweep sig-
nals.
SOURce Subsystem