Multi-Instrument Setups
R&S
®
SMW200A
609User Manual 1175.6632.02 ─ 16
Final time alignment can require adding I/Q or trigger delay. However, once synchroni-
zation is achieved, the master-slave mode ensures reproducibility of triggering, if the
used waveform, sample rate and cabling are not changed.
Test setup
Regarding the synchronization signal, the instruments can be connected (cascaded)
in a star configuration (see Figure 9-2) or in a daisy chain (see Figure 9-1).
Additionally, all instruments in the setup have to use the same reference frequency.
You can distribute the internal reference frequency of one of the instruments, regard-
less if master or slave, to all other or supply all instruments from an external reference
frequency source.
Required options
●
Option wideband baseband generator (R&S SMW-B9)
●
Option wideband baseband main module (R&S SMW-B13XT)
9.3.1.1 Multi-Instrument Settings
Option: R&S SMW-B9
Access:
► Use one of the following:
● Select "Taskbar > System Config > System Configuration > Multi Instrument"
● Select "Block Diagram > Fading > System Configuration > Multi Instrument"
● Select "Block Diagram > I/Q Stream Mapper > System Configuration > Multi
Instrument"
● On the instrument front panel, press SETUP.
Select "System Configuration > Multi Instrument"
For step-by-step description, see Chapter 9.3.1.2, "How to Connect and Configure
Instruments in Master-Slave Mode", on page 611.
Generating Time-Aligned Baseband Signals