Network Operation and Remote Control
R&S
®
SMW200A
709User Manual 1175.6632.02 ─ 16
GPIB address
The controller must address the instrument with the GPIB bus channel (see Chap-
ter 12.4.3, "GPIB Address Settings", on page 719). GPIB provides channel
addresses from 0 to 30.
The GPIB resource string syntax is as follows:
GPIB
= denotes the used interface
<channel address> = the used channel
[::INSTR] = indicates the instrument resource class (optional)
Note: If the VISA implementation supports the GPIB interface, you can optionally
define the VISA Instrument Control Resource (INSTR). It is used to define the basic
operations and attributes for a device, such as reading, writing, or triggering.
Notes and characteristics
In connection with the GPIB interface, note the following:
●
Up to 15 instruments can be connected.
●
The total cable length is restricted to a maximum of 15 m, or 2 m times the number
of devices, whichever is less. The cable length between two instruments should not
exceed 2 m.
●
A wired "OR"-connection is used if several instruments are connected in parallel,
since the slowest instrument determines the speed.
Any connected IEC bus cable must be terminated by an instrument or controller.
12.2.4 LXI Browser Interface
The LXI browser interface allows easy configuration of the LAN and remote control of
the R&S SMW without additional installation requirements. The instrument's LXI
browser interface works correctly with all W3C compliant browsers.
See Chapter 12.18.1, "LXI Functionality", on page 762 for more about LXI.
The LAN settings are configured using the instrument's LXI browser interface descri-
bed in Chapter 12.5.2.1, "LAN Configuration", on page 726. The LXI status settings
in the R&S SMW are described in Chapter 12.5.1, "LXI Status Settings",
on page 723.
12.3 Remote Control Programs and Libraries
This section shows how the remote-control programs access the instrument, and the
libraries they require for the appropriate interface protocols.
Remote Control Programs and Libraries