Remote Control Commands
R&S
®
SMW200A
1031User Manual 1175.6632.02 ─ 16
Parameters:
<State> 0 | 1 | OFF | ON
*RST: 0
Example: See Example "Create a table with user-defined correction values
for the RF level" on page 1026 .
Manual operation: See "State" on page 531
[:SOURce]:CORRection:CSET:CATalog?
Queries a list of available user correction tables.
Return values:
<Catalog> string
List of list filenames, separated by commas
Example: See Example "Create a table with user-defined correction values
for the RF level" on page 1026 .
Usage: Query only
Manual operation: See "User Cor. Data" on page 532
[:SOURce]:CORRection:CSET:DELete <Filename>
Deletes the specified user correction list file.
Setting parameters:
<Filename> string
Filename or complete file path; file extension is optional.
Example: See Example "Create a table with user-defined correction values
for the RF level" on page 1026 .
Usage: Setting only
Manual operation: See "User Cor. Data" on page 532
13.18.5.2 SGamma Correction Data
The signal generator can perform a measured-value correction taking the complex
reflection coefficient (source gamma) of the signal source into account. The state must
be activated.
[:SOURce<hw>]:CORRection:CSET:DATA:POWer:SGAMma:STATe <State>
Activates the auto fill process to create a list with SGamma correction data.
SGamma correction is possible when the table is filled with multi level user correction
data.
SOURce Subsystem