Configuring the Signal in the RF Domain
R&S
®
SMW200A
531User Manual 1175.6632.02 ─ 16
4. Select "State > On".
The "UCOR" dialog contains all settings for creating and handling files with user-
defined level correction values.
The remote commands required to define these settings are described in Chap-
ter 13.18.5, "SOURce:CORRection Subsystem", on page 1026.
Settings
State............................................................................................................................531
User Correction...........................................................................................................531
User Cor. Data............................................................................................................ 532
Edit User Cor. Data.....................................................................................................532
State
Activates user correction.
The R&S SMW displays a "UCOR" status message in the frequency and level display.
User correction cannot be activated simultaneously with absolute level compensation
based on user-defined frequency response values. These functions exclude each
other; only one of them can be used at a time.
See user manual R&S
®
SMW-K544 User-Defined Frequency Response Correction.
Remote command:
[:SOURce<hw>]:CORRection[:STATe] on page 1030
User Correction
Indicates the current value for level correction.
Remote command:
[:SOURce<hw>]:CORRection:VALue? on page 1030