General Instrument Functions
R&S
®
SMW200A
674User Manual 1175.6632.02 ─ 16
Impedance Clock/Trigger Input
Selects the input impedance for the external trigger/clock inputs.
Set the value to 1 kΩ/GND for high clock rates.
Remote command:
[:SOURce]:INPut:USER:TRIGger:IMPedance on page 862
[:SOURce]:INPut:USER:CLOCk:IMPedance on page 862
Clock Input Slope
Sets the polarity of the active slope of an externally applied clock signal.
Remote command:
[:SOURce]:INPut:USER:CLOCk:SLOPe on page 862
Trigger Input Slope
Sets the polarity of the active slope of an applied instrument trigger.
Remote command:
[:SOURce]:INPut:USER:TRIGger:SLOPe on page 863
11.2.6 Local Connectors Settings
Access:
1. Perform one of the follwoing:
In the "Block Diagram", select the T/M/C status LEDs to access the "Trigger
Marker Clock" dialog and:
Select a local T/M/C connector or
Select "Local Connector Settings"
Select "Baseband > Custom Digital Modulation > Trigger In > Local Connector
Settings" or respectively the "Trigger In", "Marker" or "Clock" tab of the corre-
sponding digital standard.
The "Local Connectors" dialog comprises the settings necessary to route the sig-
nals to the local connectors and to determine the physical characteristics of the
input connectors for trigger, clock and marker signals.
Configuring the Local and Global Connectors