Remote Control Commands
R&S
®
SMW200A
862User Manual 1175.6632.02 ─ 16
Manual operation: See "Execute Trigger" on page 673
[:SOURce]:INPut:USER:CLOCk:LEVel <Level>
[:SOURce]:INPut:USER:TRIGger:LEVel <Level>
Sets the threshold for any input signal at the USER1-3 connectors.
Parameters:
<Level> float
Range: 0.1 to 2
Increment: 0.1
*RST: 1
Example: See Example "Global connectors settings configuration"
on page 857
Manual operation: See "Threshold USER1-3 Input" on page 673
[:SOURce]:INPut:USER:PULM:LEVel <Level>
Sets the threshold for any input signal at the USER4-6 connectors.
Parameters:
<Level> float
Range: 0.1 to 2
Increment: 0.1
*RST: 1
Default unit: V
Example: See Example "Global connectors settings configuration"
on page 857
Manual operation: See "Threshold USER4-6 Input/Threshold Pulse Input"
on page 514
[:SOURce]:INPut:USER:CLOCk:IMPedance <Impedance>
[:SOURce]:INPut:USER:TRIGger:IMPedance <Impedance>
Selects the input impedance for the external trigger inputs.
Parameters:
<Impedance> G1K | G50
*RST: G1K
Example: See Example "Global connectors settings configuration"
on page 857
Manual operation: See "Impedance Clock/Trigger Input" on page 674
[:SOURce]:INPut:USER:CLOCk:SLOPe <Slope>
Sets the polarity of the active slope of an externally applied clock signal.
SOURce Subsystem