Configuring the Signal in the RF Domain
R&S
®
SMW200A
435User Manual 1175.6632.02 ─ 16
Settings
Amplitude.................................................................................................................... 435
Digital Attenuation.......................................................................................................435
Mode .......................................................................................................................... 436
Setting Characteristics ............................................................................................... 436
Attenuator Level Range ............................................................................................. 437
Readjust......................................................................................................................437
Adjust Level At Current Settings................................................................................. 438
Limit ............................................................................................................................438
Offset ..........................................................................................................................438
User Variation............................................................................................................. 438
Variation Active.............................................................................................438
Variation Step............................................................................................... 439
Amplitude
Sets the level of the RF signal.
The value is offset-free and corresponds to the level at the RF A/RF B connector.
Note: Suppressed values in the status bar
For security concerns or certain operating modes, you can hide the frequency and
level display in the status bar.
The display has been disabled for security reasons.
See:
Annotation Frequency
Annotation Amplitude
The display is disabled when:
List mode is running, see Chapter 7.6, "Varying the RF Signal in List or Sweep
Mode", on page 457.
Digital only signals are generated, see the user manual "R&S SMW-K551 Gen-
eration of Digital "Slow IQ" Signals"
See Chapter 7.3.4.1, "Displayed RF Frequency and Level Values with Downstream
Instruments", on page 440.
Remote command:
[:SOURce<hw>]:POWer:POWer on page 1099
Note: The SCPI command [:SOURce<hw>]:POWer[:LEVel][:IMMediate][:
AMPLitude] sets the level of the "Level" display.
This means, the level containing offset.
Digital Attenuation
Attenuates or raises the level of the internal, digitally modulated I/Q baseband signals,
and thus the level of the RF signal accordingly.
The function allows fast level changes of the internal I/Q signals. You can determine an
attenuation value between -3 dB and +80 dB. The instrument calculates the digital
attenuation to the Level
PEP
and the Level
rms
as follows:
Configuring RF Frequency and Level