Reference Information for Remote Control
R&S
®
SMW200A
1192User Manual 1175.6632.02 ─ 16
"SCPI confirmed" in the command reference chapters. Commands without this
SCPI label are device-specific; however, their syntax follows SCPI rules as per-
mitted by the standard.
Instrument responses
Instrument responses (response messages and service requests) are messages the
instrument sends to the controller after a query. They can contain measurement
results, instrument settings and information on the instrument status.
B.1.2 LAN Interface Messages
In the LAN connection, the interface messages are called low–level control messages.
These messages can be used to emulate interface messages of the GPIB bus.
Command Long term Effect on the instrument
&ABO Abort Aborts processing of the commands just received.
&DCL Device Clear Aborts processing of the commands just received and
sets the command processing software to a defined initial
state. Does not change the instrument setting.
&GTL Go to Local Transition to the "local" state (manual control). (The
instrument automatically returns to remote state when a
remote command is sent UNLESS &NREN was sent
before.)
&GTR Go to Remote Enables automatic transition from local state to remote
state by a subsequent remote command (after &NREN
was sent).
&GET Group Execute Trigger Triggers a previously active instrument function (e.g. a
sweep). The effect of the command is the same as with
that of a pulse at the external trigger signal input.
&LLO Local Lockout Disables transition from remote control to manual control
by means of the front panel keys.
&NREN Not Remote Enable Disables automatic transition from local state to remote
state by a subsequent remote command. (To re-activate
automatic transition use &GTR.)
&POL Serial Poll Starts a serial poll.
B.1.3 SCPI Command Structure
SCPI commands consist of a header and, in most cases, one or more parameters. The
header and the parameters are separated by a "white space" (ASCII code 0 to 9, 11 to
32 decimal, e.g. blank). The headers may consist of several mnemonics (keywords).
Queries are formed by appending a question mark directly to the header.
The commands can be either device-specific or device-independent (common com-
mands). Common and device-specific commands differ in their syntax.
Additional Basics on Remote Control