Loading...
Maintenance
R&S
®
SMW200A
1156User Manual 1175.6632.02 ─ 16
14.3.2.1 Check Front Panel Settings
Access:
Select "System Config > Setup > Maintenance > Check Front Panel".
Reflecting the front panel, the "Check Front Panel" dialog contains all functions to
test the operating elements of the instrument. In addition, you can check the touch-
sensitive functionality by dragging one or more lines across the screen with your
finger.
Calibrate
Executes the touchscreen calibration.
See Chapter 2.1.5.5, "Calibrating the Touchscreen", on page 51 for executing the test.
Touchscreen Test
Enables you to perform a touchscreen test. This test checks whether the touchscreen
area has gaps in the response.
See "How to perform the touchscreen test" on page 1159
14.3.2.2 How to Perform the Front Panel Tests
How to perform the touchscreen calibration
The screen responds to the touch of your finger, that means you can select or activate
an item by tapping it. Due to the design of the touchscreen, the capture area is firmly
set to a certain size and you do not need to define a range. But if the instrument does
not respond correctly to the touch, we recommend that you calibrate the touchscreen.
Performing Maintenance Tasks

Table of Contents