Remote Control Commands
R&S
®
SMW200A
1070User Manual 1175.6632.02 ─ 16
Parameters:
<Period> float
Range: 1E-6 to 100
Increment: 10E-9
*RST: 0.001
Example: See Example "Configuring the LF generator" on page 1062.
Manual operation: See "Period" on page 504
[:SOURce<hw>]:LFOutput<ch>:SHAPe:TRIangle:RISE <Rise>
Selects the rise time for the triangle single of the LF generator.
Parameters:
<Rise> float
Range: 1E-6 to 100
Increment: 10E-9
*RST: 0.5E-3
Example: See Example "Configuring the LF generator" on page 1062.
Manual operation: See "Triangle Rise" on page 504
13.18.10.2 LF Sweep Settings
With the commands described in this section, you can configure the sweep of the LF
signal.
[:SOURce<hw>]:LFOutput:SWEep[:FREQuency]:DWELl................................................. 1070
[:SOURce<hw>]:LFOutput:SWEep[:FREQuency]:EXECute..............................................1071
[:SOURce<hw>]:LFOutput:SWEep[:FREQuency]:MODE..................................................1071
[:SOURce<hw>]:LFOutput:SWEep[:FREQuency]:POINts.................................................1071
[:SOURce<hw>]:LFOutput:SWEep[:FREQuency]:RETRace............................................. 1072
[:SOURce<hw>]:LFOutput:SWEep[:FREQuency]:RUNNing?............................................1072
[:SOURce<hw>]:LFOutput:SWEep[:FREQuency]:SHAPe.................................................1072
[:SOURce<hw>]:LFOutput:SWEep[:FREQuency]:SPACing.............................................. 1072
[:SOURce<hw>]:LFOutput:SWEep[:FREQuency]:STEP:LOGarithmic................................1073
[:SOURce<hw>]:LFOutput:SWEep[:FREQuency]:STEP[:LINear]...................................... 1073
[:SOURce<hw>]:LFOutput:SWEep[:FREQuency]:DWELl <Dwell>
Sets the dwell time for each frequency step of the sweep.
Parameters:
<Dwell> float
Range: 1E-3 to 100
Increment: 100E-6
*RST: 15E-3
Default unit: s
Example: See Example "Setup an LF sweep" on page 1061.