Remote Control Commands
R&S
®
SMW200A
961User Manual 1175.6632.02 ─ 16
Clock Commands
[:SOURce<hw>]:BB:ARBitrary:CLOCk:SOURce <Source>
Selects the clock source:
INTernal: Internal clock reference
ELCLock: External local clock
EXTernal = ELCLock: Setting only
Provided for backward compatibility with other Rohde & Schwarz signal generators
Parameters:
<Source> INTernal | ELCLock | EXTernal
*RST: INTernal
Example: see Example "Managing waveform files" on page 920
Options: ELCLock requires R&S SMW-B10
Manual operation: See "Clock Source" on page 260
[:SOURce<hw>]:BB:ARBitrary:CLOCk:MODE <Mode>
Enters the type of externally supplied clock.
Parameters:
<Mode> SAMPle
*RST: SAMPle
Example: See Example "Managing waveform files" on page 920
Options: R&S SMW-B10
Manual operation: See "Clock Mode" on page 261
13.18.4.4 Waveform, Data and List Format
The R&S SMW uses a simple tag-oriented format for externally or internally generated
waveforms, data and control lists. Files with this format can be transmitted via the
GPIB interface from an external computer to the instrument and vice versa. These files
are transmitted as a binary data block, using SCPI command [:SOURce<hw>]:BB:
ARBitrary:WAVeform:DATA, [:SOURce<hw>]:BB:DM:DLISt:DATA, and [:
SOURce<hw>]:BB:DM:CLISt:DATA.
Tag general format
Tags are self-contained information units, enclosed in braces { }. Their general format
is {Name: Data} or {Name-Length: Data}. The colon separates the name part
and the data part. The colon can be followed by a space for the sake of legibility.
The Name identifies the tag. It is always expressed in capital letters.
The Data is tag-specific, and usually it is in ASCII plain text.
SOURce Subsystem