Remote Control Commands
R&S
®
SMW200A
1033User Manual 1175.6632.02 ─ 16
[:SOURce<hw>]:CORRection:CSET:DATA:SGAMma:PHASe <Phase1>[,
<PhaseN>]
Sets the phase angle of the reflection coefficient of the DUT for SGamma correction
(multi level user correction data).
Parameters:
<Phase1> float
Range: -360 to 360
Increment: 0.1
<PhaseN> float
Range: -360 to 360
Increment: 0.1
Example:
CORR:CSET:DATA:POW:SGAM:PHAS
-360.0,-180.0,180.0,360.0
Sets the phase values.
[:SOURce<hw>]:CORRection:CSET:DATA:SGAMma:PHASe:POINts?
Queries the number of phase entries in the selected multi level user correction table.
Return values:
<Points> integer
Example:
CORR:CSET "/var/user/temp/multilevelucor"
CORR:CSET:DATA:SGAM:PHAS:POIN?
// 40
// the table contains 40 phase values
Usage: Query only
13.18.5.3 Correction Data Exchange
With the following commands, you can configure user correction lists and export or
import them accordingly.
[:SOURce<hw>]:CORRection:DEXChange:AFILe:CATalog?
Queries the available ASCII files for export or import of user correction data in the cur-
rent or specified directory.
Return values:
<Catalog> string
List of ASCII files *.txt or *.csv, separated by commas.
Example: See Example "Create a table with user-defined correction values
for the RF level" on page 1026 .
Usage: Query only
Manual operation: See "Select Source/Select Destination" on page 491