Remote Control Commands
R&S
®
SMW200A
1073User Manual 1175.6632.02 ─ 16
Parameters:
<Spacing> LINear | LOGarithmic
*RST: LINear
Example: See Example "Setup an LF sweep" on page 1061.
Manual operation: See " Spacing" on page 479
[:SOURce<hw>]:LFOutput:SWEep[:FREQuency]:STEP:LOGarithmic
<Logarithmic>
Sets the step width factor for logarithmic sweeps to calculate the frequencies of the
steps.
For information on how the value is calculated and the interdependency with other
parameters, see Chapter 7.6.2.1, "Correlating Parameters in Sweep Mode",
on page 467
Parameters:
<Logarithmic> float
The unit is mandatory
Range: 0.01 to 100
Increment: 0.01
*RST: 1
Default unit: PCT
Example: See Example "Setup an LF sweep" on page 1061.
Manual operation: See " Step Linear/Step Logarithmic " on page 480
[:SOURce<hw>]:LFOutput:SWEep[:FREQuency]:STEP[:LINear] <Linear>
Sets the step width for the linear sweep.
For information on how the value is calculated and the interdependency with other
parameters, see Chapter 7.6.2.1, "Correlating Parameters in Sweep Mode",
on page 467
Parameters:
<Linear> float
Range: 0 to STOP-STARt
Increment: 0.1
*RST: 1000
Example: See Example "Setup an LF sweep" on page 1061.
Manual operation: See " Step Linear/Step Logarithmic " on page 480
13.18.11 SOURce:LIST Subsystem
The SOURce:LIST subsystem contains all commands for defining lists and for han-
dling of list files.
SOURce Subsystem