Configuring the Internal Baseband Source
R&S
®
SMW200A
236User Manual 1175.6632.02 ─ 16
Example:
In the following example, the marker signal is defined by a pattern 100100....
Marker mode ON/OFF ratio
Similar to "Pulse" but with independent lengths of the On and Off periods. The length
of the periods is entered as a number of symbols/sample periods.
Example:
In the following example, the marker signal is defined by an "On Time" = 1 and "Off
Time" = 2.
Marker mode CList
This is not one of the most common modes.
The instrument generates a marker signal that is defined in the selected control list.
The R&S SMW provides a graphical interface for convenient definition of control sig-
nals among others also for the marker signals.
Marker Minimum Duration
The period length of several marker signals is specified as a number of symbols/
sample, in one of the two ways: directly or as mathematical function, where the marker
period length is derived from other marker parameters. (See "Marker Modes"
on page 235)
Irrespectively of the way it is defined, if the Wideband baseband is used, the duration
of the shortest marker period length is a value that depends on used sample or symbol
rate. The possible sample or symbol rates in turn depend on the installed options, see
Table 4-7.
Table 4-7: Minimum marker duration if the wideband generator is used
Sample rate or symbol rate Minimum marker duration,
Samples/symbols
Required options
300 MHz 1 R&S SMW-B9
300 MHz to 600 MHz 2 R&S SMW-B9