Configuring the Internal Baseband Source
R&S
®
SMW200A
376User Manual 1175.6632.02 ─ 16
"Chirp"
Rapid crest factor optimization regardless of the number of carriers.
The minimal crest factor is 3 dB. This value is obtained, if the follow-
ing prerequisites are fulfilled:
All carriers in the multicarrier signal are switched on.
All carriers have identical power.
In a configuration which differs from this, the achievable crest factor is
worse.
"Target Crest"
Optimization of the crest factor to a desired value for all carrier config-
urations.
The optimization time depends on the number of carriers and the
desired crest factor. Computation time increases only when the num-
ber of carriers exceeds 256 and the crest factor is below 4 dB. The
desired value can be entered in "Desired Crest Factor".
Note: Optimization can be canceled at any time, and the current
value being displayed at that moment is then used.
Remote command:
[:SOURce<hw>]:BB:MCCW:CFACtor:MODE on page 990
Desired Crest Factor
For optimization "Target Crest", defines the desired crest factor.
Remote command:
[:SOURce<hw>]:BB:MCCW:CFACtor on page 990