Network Operation and Remote Control
R&S
®
SMW200A
742User Manual 1175.6632.02 ─ 16
4. In the toolbar bar of the "SCPI Remote Trace" page, select "live mode > on" and
"logging > on".
"live mode > on" displays all commands and responses, and "logging > on" also
traces messages.
If you now control the R&S SMW with SCPI commands, using an appropriate tool,
the LXI function records the information sent and received.
The function records all sent commands, received responses and messages, and
saves them in an internal database. If "live mode" is disabled, you can display the
recent traces upon request, using the "refresh" button. You can also save the log in
a file.
Note: The diagnostics functionality will be extended in later releases, e.g. to down-
load or upload SCPI command files from / to the instrument.
12.10 How to Return to Manual Operation
Before returning to manual control, command processing must be completed. Other-
wise, the instrument switches back to remote control immediately.
1. To return from "Remote" state to manual state, perform one of the following:
● On the controller, use the command >L
Note: If &NREN has been set before >L is locked. Use >R instead.
● In the status bar, select the "Remote" icon.
● On the front panel, press the LOCAL key.
● In the block diagram, select "Context sensitive menu > Key Emulation > Local"
How to Return to Manual Operation