Remote Control Commands
R&S
®
SMW200A
900User Manual 1175.6632.02 ─ 16
[:SOURce<hw>]:DM:FILTer:PARameter <Parameter>
Sets the filter parameter of the currently selected filter type.
To set the filter type, use command [:SOURce<hw>]:BB:DM:FILTer:TYPE
on page 898.
Parameters:
<Parameter> float
Range: 0.05 to 2.5
Increment: 0.01
*RST: 0.35
Example: see [:SOURce<hw>]:BB:DM:FILTer:PARameter:SPHase
on page 899
Modulation and Coding Settings
[:SOURce<hw>]:BB:DM:CODing <Coding>
Selects the modulation coding.
Parameters:
<Coding> OFF | DIFF | DPHS | DGRay | GRAY | GSM | NADC | PDC |
PHS | TETRa | APCO25 | PWT | TFTS | INMarsat | VDL |
EDGE | APCO25FSK | ICO | CDMA2000 | WCDMA |
APCO258PSK
OFF
The coding is automatically disabled if the selected modulation
type is not possible with the coding that has been set
DPHS
Phase Difference
DGRay
Difference + Gray
*RST:
GSM
Example: See Example "Performing general tasks" on page 891
Options: DPHS|GSM|INMarsat require R&S SMW-B10
Manual operation: See "Coding" on page 267
[:SOURce<hw>]:BB:DM:FORMat <Format>
Sets the modulation type.
When a standard is selected ([:SOURce<hw>]:BB:DM:STANdard), the modulation
type is set to the default value.
SOURce Subsystem