Elements and Controls of the Graphical User Interface
R&S
®
SMW200A
1230User Manual 1175.6632.02 ─ 16
ListMode
List mode is active.
The values of the frequency/level pairs in the selected list are set for the chosen dwell
time.
AttFixed
Attenuator fixed mode is active.
The uninterrupted level settings are made in a fixed range without attenuator switching.
The variation range is set automatically when this mode is activated. The range is dis-
played under "Attenuator Fixed Range" in the "Level" dialog.
UCorr
User Correction is active.
The level is corrected by the given values in the selected user correction list. Correc-
tion is performed by the user-defined list values being added to the output level for the
respective RF frequency. With frequencies which are not contained in the list, the level
correction is determined by interpolation of the closest correction values.
OvenCold
The reference oscillator has not yet reached its nominal frequency.
When switching on from the STANDBY mode, the specified frequency accuracy is
reached immediately. If the power switch was switched off, the reference oscillator
needs some warm-up time to reach its nominal frequency. During this period of time,
the output frequency does not yet reach its final value either.
D.5 Elements Displayed for Interactions
The signal generator is primarily configured by settings dialogs, but also by active ele-
ments in the block diagram and the status bar. Additional interactive elements such as
tooltips or on-screen keypads help you to conveniently perform the settings.
Dialogs
It is distinguished between settings dialogs and diagrams.
Basically, a dialog has a parent name corresponding to its content.
Elements Displayed for Interactions