Remote Control Commands
R&S
®
SMW200A
1048User Manual 1175.6632.02 ─ 16
[:SOURce<hw>]:INPut:MODext:IMPedance<ch> <Impedance>
Sets the impedance (50 kOhm or High = 100 kOhm to ground) for the externally sup-
plied modulation signal.
Parameters:
<Impedance> G50 | HIGH
*RST: HIGH
Example:
INP:MOD:IMP 50
sets 50 kOhm to ground.
Manual operation: See "Impedance" on page 505
[:SOURce<hw>]:INPut:TRIGger:SLOPe <Slope>
Sets the polarity of the active slope of an applied instrument trigger.
Parameters:
<Slope> NEGative | POSitive
*RST: POSitive
Manual operation: See "Trigger Slope" on page 480
13.18.8 SOURce:IQ Subsystem
The SOURce:IQ subsystem contains the commands for configuring the I/Q modula-
tion.
Required options
See Chapter 6.1, "Required Options", on page 415.
[:SOURce<hw>]:IQ:SOURce.........................................................................................1048
[:SOURce<hw>]:IQ:STATe............................................................................................1049
[:SOURce<hw>]:IQ:GAIN..............................................................................................1049
[:SOURce<hw>]:BB:IQGain.......................................................................................... 1049
[:SOURce<hw>]:IQ:CREStfactor....................................................................................1050
[:SOURce<hw>]:IQ:SWAP[:STATe]............................................................................... 1050
[:SOURce<hw>]:IQ:WBSTate........................................................................................1051
[:SOURce<hw>]:IQ:SOURce <Source>
Selects the input signal source for the I/Q modulator.
SOURce Subsystem