EasyManua.ls Logo

Rohde & Schwarz SMW200A - Data List Editor

Rohde & Schwarz SMW200A
1312 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Configuring the Internal Baseband Source
R&S
®
SMW200A
280User Manual 1175.6632.02 ─ 16
Remote command:
[:SOURce<hw>]:BB:DM:PRAMp:ATTenuation on page 904
In Baseband Only
Restricts power ramping to the baseband signal.
"Off"
Level attenuation is affected via the attenuator stages in the RF sec-
tion; only the remaining part is attenuated in the baseband. The sig-
nal is issued at the RF output with the defined level values.
"On"
Level attenuation is affected in the baseband only.
Note:
This setting is mandatory in the following cases:
When only the baseband signal is issued at the I/Q outputs
It is thus ensured that, with power ramping active, this signal is
output with the defined level values.
In case of two-path instruments, when one baseband signal is
applied to two RF paths.
Remote command:
[:SOURce<hw>]:BB:DM:PRAMp:BBONly[:STATe] on page 904
4.6.3.7 Data List Editor
The R&S SMW provides the following ways to create a data list file:
To use the dedicated "Data List Editor" and create a file with extension *.dm_iqd,
see "To create data lists manually" on page 288
To use the tag-oriented format and create a data list file, see "To create a data list
file using tag file format" on page 981
To use SCPI commands and create a file in binary format, see "To create a data
list in binary format" on page 982
To access the "Data List Editor" dialog:
1. Select "Baseband > Custom Dig Mod > Data Source > Data List"
2. Select "Select Data List"
3. Navigate to the required directory.
Select "New" and enter a filename.
Generating Custom Digital Modulated Signals

Table of Contents