Remote Control Commands
R&S
®
SMW200A
904User Manual 1175.6632.02 ─ 16
Parameters:
<RDelay> float
Range: 0 to 4
Increment: 0.01
*RST: 0
Default unit: symbol
Example: See Example "Enable power ramping" on page 893
Options: R&S SMW-B10
Manual operation: See "Rise Delay" on page 279
[:SOURce<hw>]:BB:DM:PRAMp:ATTenuation
<Attenuation>
Sets the level attenuation for signal ranges that are flagged with level attribute attenu-
ated by the control signal.
Parameters:
<Attenuation> float
Range: 0 to 50
Increment: 0.1
*RST: 15
Default unit: dB
Example: See Example "Enable power ramping" on page 893
Options: R&S SMW-B10
Manual operation: See "Attenuation" on page 279
[:SOURce<hw>]:BB:DM:PRAMp:BBONly[:STATe] <State>
Enables power ramping in the baseband only or mixed power ramping in the baseband
and the RF section.
The ON setting is mandatory if, with power ramping active, only the baseband signal is
output (I/Q outputs)
In case of two-path instruments, the ON setting is also mandatory if a baseband signal
is applied to two RF paths (RF A and RF B).
Only then can a signal with a defined, predictable level be output.
Parameters:
<State> 0 | 1 | OFF | ON
*RST: 0
Example: See Example "Enable power ramping" on page 893
Options: R&S SMW-B10
Manual operation: See "In Baseband Only" on page 280