File and Data Management
R&S
®
SMW200A
627User Manual 1175.6632.02 ─ 16
10.2.4 Reference
See Table 10-1 for an overview of the main generator settings that are affected by the
corresponding preset functions. While the regular PRESET key primarily resets the sig-
nal relevant parameters of the instrument, the "Factory Preset" affects almost all instru-
ment settings.
For information on the default values of further parameters, refer to the description of
the corresponding remote commands.
Table 10-1: Key parameters affected by preset and factory preset
Parameter Preset value Preset Factory
Preset
RF frequency 1 GHz x x
RF level (RF output) off x x
RF OFF mode - - x
Offsets 0 x x
Modulation state off x x
Uninterrupted level settings off x x
Level attenuator mode auto x x
Level ALC (internal level control) auto x x
Level UCOR (user correction) off x x
LF output state off x x
Sweep state off x x
List mode state off x x
Reference frequency settings (reference oscillator) - - x
Power on settings (Level/EMF) - - x
Network settings - - x
Hostname
1)
- - x
GPIB address - - x
Start/Stop display update - - x
Display and keyboard settings - - x
Preset behavior: Keep connections to external instru-
ments
off - x
Password and settings protected by passwords (e.g. dis-
abled LAN or USB)
- - -
Security settings - - -
User files (setups
2)
, data lists, etc.)
- - -
Restoring the (Default) Instrument Configuration