Signal Routing and System Configuration
R&S
®
SMW200A
196User Manual 1175.6632.02 ─ 16
5. Select "RF A > On" and "RF B > On".
3.9.4 How to Route the I/Q Streams to the Output Interfaces
This example shows how to enable the instrument to output the EUTRA/LTE 1x4x4
MIMO signal at both RF outputs and both I/Q analog outputs.
1. Select "Block Diagram > I/Q Stream Mapper".
2. In the stream to connector mapping matrix, tap on a crossing cell to enable the
routing of the selected stream to the corresponding connector.
For example, enable "Stream A > RF A", "Stream B > RF B", "Stream C > I/Q OUT
1" and "Stream D > I/Q OUT 2".
How to Connect External Instruments and Configure the Signal Flow