Remote Control Commands
R&S
®
SMW200A
825User Manual 1175.6632.02 ─ 16
Example:
SENS1:TYPE?
Queries the type of sensor.
Response: NRP-Z21
The R&S NRP-Z21 sensor is used.
Usage: Query only
Manual operation: See " Sensor type and serial number " on page 547
:SENSe<ch>[:POWer]:ZERO
Performs zeroing of the sensor.
Zeroing is required after warm-up, i.e. after connecting the sensor.
Note: Switch off or disconnect the RF power source from the sensor before zeroing.
We recommend that you zero in regular intervals (at least once a day), if:
●
The temperature has varied more than about 5 °C.
●
The sensor has been replaced.
●
You want to measure very low power.
Example:
SENS1:ZERO
Executes zeroing.
Usage: Event
Manual operation: See " Zero " on page 547
13.17 SCONfiguration Subsystem
The SCONfiguration subsystem contains the commands for defining the system
configuration setting, like stream mapping and used fading configuration.
Required options
See "Required options" on page 138.
Suffixes
The following suffixes are used:
Suffix Value range Description
STReam<st>
1 to 8 Available IQ streams
CODer<ch>,
BBMM<ch>,
IQOutput<ch>
1 to 2 DIG IQ connector
FADer<ch>
1 to 4 DIG IQ connector
SCONfiguration Subsystem