Remote Control Commands
R&S
®
SMW200A
824User Manual 1175.6632.02 ─ 16
:SENSe<ch>[:POWer]:SOURce <Source>
Determines the signal to be measured.
Note: When measuring the RF signal, the sensor considers the corresponding correc-
tion factor at that frequency, and uses the level setting of the instrument as reference
level.
Parameters:
<Source> A | USER | RF | B
*RST: A
Example: See Example "Performing a power measurement with a fixed fil-
ter" on page 814.
Manual operation: See " Use Frequency Of" on page 548
:SENSe<ch>[:POWer]:STATus[:DEVice]?
Queries if a sensor is connected to the instrument.
Return values:
<Status> 0 | 1 | OFF | ON
*RST: 0
Example:
SENS1:STAT?
Response: 1
A sensor is connected.
Usage: Query only
Manual operation: See " State " on page 547
:SENSe<ch>[:POWer]:SVERsion?
Queries the software version of the connected R&S NRP power sensor.
Return values:
<SVersion> string
Example:
SENS1:POW:SVER?
Queries the software version of the power sensor.
Usage: Query only
Manual operation: See "Current Sensors" on page 551
:SENSe<ch>[:POWer]:TYPE?
Queries the sensor type. The type is automatically detected.
Return values:
<Type> string
SENSe, READ, INITiate and SLISt Subsystems