Remote Control Commands
R&S
®
SMW200A
823User Manual 1175.6632.02 ─ 16
Example:
SENS:LOGG:STAT ON
Activates recording of the power measurement of the first sen-
sor.
Manual operation: See "Enable Logging" on page 550
:SENSe<ch>[:POWer]:OFFSet <Offset>
Sets a level offset which is added to the measured level value after activation with
command :SENSe<ch>[:POWer]:OFFSet:STATe. The level offset allows, e.g. to
consider an attenuator in the signal path.
Parameters:
<Offset> float
Range: -100.0 to 100.0
*RST: 0
Default unit: dB
Example:
SENS1:POW:OFFS 10.0
Sets a level offset of 10 dB
Manual operation: See " Level Offset State,Level Offset" on page 548
:SENSe<ch>[:POWer]:OFFSet:STATe <State>
Activates the addition of the level offset to the measured value. The level offset value is
set with command :SENSe<ch>[:POWer]:OFFSet.
Parameters:
<State> 0 | 1 | OFF | ON
*RST: 0
Example:
SENS1:POW:OFFS 0.4dB
Sets a level offset of 0.4 dB
SENS1:POW:OFFS:STAT ON
A level offset of 0.4 dB is added to the measured value.
Manual operation: See " Level Offset State,Level Offset" on page 548
:SENSe<ch>[:POWer]:SNUMber?
Queries the serial number of the sensor.
Return values:
<SNumber> string
Example:
SENS1:SNUM?
Queries the serial number.
Usage: Query only
Manual operation: See " Sensor type and serial number " on page 547