Configuring the Signal in the RF Domain
R&S
®
SMW200A
483User Manual 1175.6632.02 ─ 16
Remote command: 
[:SOURce<hw>]:LIST:RESet on page 1085
Dwell Time Mode
Selects either variable dwell times or a fixed dwell time to perform signal generation in
list mode.
"Global"
Uses the same dwell time for all lists steps, set with Global Dwell
Time.
"From List"
Uses the dwell times from the list.
You can define the dwell time for each frequency/level value pair indi-
vidually, see Chapter 7.6.7, "List Editor", on page 486.
Remote command: 
[:SOURce<hw>]:LIST:DWELl:MODE on page 1078
Global Dwell Time
Sets the dwell time for Dwell Time Mode > "Global".
See also "Dwell time" on page 472.
Remote command: 
[:SOURce<hw>]:LIST:DWELl on page 1078
Current Index
Sets the list index for list processing in "Step" mode. In the other modes, the index indi-
cates the current step.
Remote command: 
[:SOURce<hw>]:LIST:INDex on page 1080
Trigger Slope
Selects the polarity of the active slope of an applied instrument trigger.
Trigger signal is expected at the INST TRIG X connector.
"Positive"
Activates the rising edge of the trigger signal.
"Negative"
Activates the falling edge of the trigger signal.
Remote command: 
[:SOURce<hw>]:INPut:TRIGger:SLOPe on page 1048
7.6.6.2 List Mode Data Settings
Access:
1. Select "RF" > "Sweep/List" > "List mode".
 
Varying the RF Signal in List or Sweep Mode