EasyManua.ls Logo

Rohde & Schwarz SMW200A - Page 497

Rohde & Schwarz SMW200A
1312 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Configuring the Signal in the RF Domain
R&S
®
SMW200A
482User Manual 1175.6632.02 ─ 16
Settings
State ...........................................................................................................................482
Mode........................................................................................................................... 482
Execute Single............................................................................................................ 482
Reset ..........................................................................................................................482
Dwell Time Mode........................................................................................................ 483
Global Dwell Time.......................................................................................................483
Current Index ............................................................................................................. 483
Trigger Slope...............................................................................................................483
State
Activates the list mode and processes the currently selected list.
Note: Active sweep mode deactivates other sweeps or lists and vice versa.
Remote command:
[:SOURce<hw>]:FREQuency:MODE on page 1041
Mode
Selects the mode for list processing.
For detailed information on the sweep modes and the triggering, see Chapter 7.6.1,
"Signal Generation and Triggering in the Sweep and List Modes", on page 459.
"Auto"
Generates the signal by processing the frequency/level value pairs of
the list from the beginning to the end.
The list steps are performed automatically, controlled by the dwell
time. If you switch from any mode to "Auto", signal generation always
starts at the top of the list.
"Single / Extern Single"
Generates the signal by processing the frequency/level value pairs of
the list once from the beginning to the end after a trigger event.
The list steps are performed automatically, controlled by the dwell
time. If one cycle is completed, the instrument waits for the next trig-
ger event.
"Step / Extern Step"
Generates the signal by processing the frequency/level value pairs of
the list step-by-step, manually triggered.
Remote command:
[:SOURce<hw>]:LIST:TRIGger:SOURce on page 1082
[:SOURce<hw>]:LIST:MODE on page 1081
Execute Single
Manually starts list processing in "Single" mode.
Remote command:
[:SOURce<hw>]:LIST:TRIGger:EXECute on page 1082
Reset
Resets the list to the starting point.
Varying the RF Signal in List or Sweep Mode

Table of Contents