EasyManua.ls Logo

Rohde & Schwarz SMW200A - Required Options

Rohde & Schwarz SMW200A
1312 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Configuring the Internal Baseband Source
R&S
®
SMW200A
373User Manual 1175.6632.02 ─ 16
4.10.1 Required Options
The equipment layout for processing of waveform files includes
Option Standard or Wideband Baseband Generator (R&S SMW-B10/-B9) per sig-
nal path
Option Baseband main module, one/two I/Q paths to RF (R&S SMW-B13/-B13T) or
option Wideband baseband main module two I/Q paths to RF (R&S SMW-B13XT)
Option Multi-Carrier CW (R&S SMW-K61) per signal path
4.10.2 About the MCCW
The R&S SMW can calculate and generate a multi-carrier continuous wave signal build
form up to 8192 unmodulated carriers. The carrier offset and the power level of the car-
riers are user-definable. The provided settings allow individual configuration of each
carrier and the joint configuration of subset of carriers. Automatic start phase setting is
provided to minimize the crest factor.
The multi-carrier CW signals are not generated in realtime but rather computed in arbi-
trary waveform mode. Changes in the carrier settings are not considered until the
instrument is triggered to accept and apply them.
The common application fields of the multi-carrier CW signals are the receiver tests
with broadband test signals.
4.10.3 MCCW Settings
Access:
Select "Baseband > Multi-Carrier CW".
The dialog enables direct carrier configuration, access to the carrier table and the
graphical representation of the configured carriers.
Generating Multi-Carrier Continuous Wave Signal

Table of Contents