Configuring the Internal Baseband Source
R&S
®
SMW200A
261User Manual 1175.6632.02 ─ 16
Remote command:
[:SOURce<hw>]:BB:DM:CLOCk:SOURce on page 912
[:SOURce<hw>]:BB:ARBitrary:CLOCk:SOURce on page 961
[:SOURce<hw>]:BB:MCCW:CLOCk:SOURce on page 1001
Clock Mode
(requires R&S SMW-B10)
Sets the type of externally supplied clock.
Remote command:
[:SOURce<hw>]:BB:DM:CLOCk:MODE on page 912
[:SOURce<hw>]:BB:ARBitrary:CLOCk:MODE on page 961
[:SOURce<hw>]:BB:MCCW:CLOCk:MODE on page 1001
Measured External Clock
(requires R&S SMW-B10)
Provided for permanent monitoring of the enabled and externally supplied clock signal.
Remote command:
:CLOCk:INPut:FREQuency? on page 794
4.5.2.4 Local and Global Connector Settings
Each of the "Trigger In", "Marker" and "Clock" dialogs as well as the "Trigger Marker
Clock" dialog provides a quick access to the related local and global connector set-
tings.
For more information, refer to the description R&S SMW User Manual, section "Local
and Global Connectors".
4.5.3 How to Use the Trigger Settings
In test setups that combine the signal of two or more instruments and/or basebands it
is often required that:
●
The signal generation starts at a defined moment
●
The signal generation starts simultaneously (or with an exactly defined delay) in:
– All involved instruments
– All basebands of the same R&S SMW
This section explains how to fulfill the last requirement with the provided trigger set-
tings.
See also:
●
"Receiving and Providing Trigger Signals" on page 244 for background information
●
Chapter 9.3, "Generating Time-Aligned Baseband Signals", on page 608 for infor-
mation on simultaneous signal generation start in multiple instruments.
Common Functions and Settings in the Baseband Domain