Remote Control Commands
R&S
®
SMW200A
1032User Manual 1175.6632.02 ─ 16
Parameters:
<State> 0 | 1 | OFF | ON
*RST: 0
Example:
CORR:CSET:DATA:POW:SGAM:STAT ON
Activates the acquisition of multi level gamma correction data
with a connected power sensor.
[:SOURce<hw>]:CORRection:CSET:DATA:SGAMma:DELete
Deletes SGamma correction values in the selected table.
Example:
CORR:CSET:DATA:SGAM:DEL
"/var/user/temp/multilevelucor1"
Usage: Event
[:SOURce<hw>]:CORRection:CSET:DATA:SGAMma:MAGNitude <Magnitude1>[,
<MagnitudeN>]
Sets the magnitude of the reflection coefficient of the DUT for SGamma correction
(multi level user correction data).
Parameters:
<Magnitude1> float
Range: 0.0 to 1.0
Increment: 0.01
<MagnitudeN> float
Range: 0.0 to 1.0
Increment: 0.01
Example:
CORR:CSET:DATA:POW:SGAM:MAG 0.1,0.2,0.3,0.4
Sets the magnitude values.
[:SOURce<hw>]:CORRection:CSET:DATA:SGAMma:MAGNitude:POINts?
Queries the number of magnitude entries in the selected multi level user correction
table.
Return values:
<Points> integer
Example:
CORR:CSET "/var/user/temp/multilevelucor"
CORR:CSET:DATA:SGAM:MAGN:POIN?
// 40
// the table contains 40 magnitude values
Usage: Query only
SOURce Subsystem