Remote Control Commands
R&S
®
SMW200A
1069User Manual 1175.6632.02 ─ 16
Example: See Example "Configuring the LF generator" on page 1062.
Manual operation: See "Trapezoid Rise / Fall" on page 505
[:SOURce<hw>]:LFOutput<ch>:SHAPe:TRAPeze:HIGH <High>
Sets the high time for the trapezoid signal of the LF generator.
Parameters:
<High> float
Range: 1E-6 to 100
Increment: 10E-9
*RST: 250E-6
Example: See Example "Configuring the LF generator" on page 1062.
Manual operation: See "Trapezoid High" on page 505
[:SOURce<hw>]:LFOutput<ch>:SHAPe:TRAPeze:PERiod
<Period>
Sets the period of the generated trapezoid shape. The period determines the repetition
frequency of the internal signal.
Parameters:
<Period> float
Range: 1E-6 to 100
Increment: 1E-8
*RST: 1E-3
Example: See Example "Configuring the LF generator" on page 1062.
Manual operation: See "Period" on page 504
[:SOURce<hw>]:LFOutput<ch>:SHAPe:TRAPeze:RISE <Rise>
Selects the rise time for the trapezoid shape of the LF generator.
Parameters:
<Rise> float
Range: 1E-6 to 100
Increment: 10E-9
*RST: 250E-6
Example: See Example "Configuring the LF generator" on page 1062.
Manual operation: See "Trapezoid Rise / Fall" on page 505
[:SOURce<hw>]:LFOutput<ch>:SHAPe:TRIangle:PERiod <Period>
Sets the period of the generated pulse. The period determines the repetition frequency
of the internal signal.