Remote Control Commands
R&S
®
SMW200A
911User Manual 1175.6632.02 ─ 16
Example: See Example "Adjusting clock, marker and trigger settings"
on page 892
Manual operation: See "Marker Mode" on page 259
[:SOURce<hw>]:BB:DM:TRIGger:OUTPut<ch>:PATTern <Pattern>
Defines the bit pattern used to generate the marker signal.
Parameters:
<Pattern> integer
0 = marker off, 1 = marker on
Range: #B0,1 to #B111...1,32
*RST: #B,1
Example: See Example "Adjusting clock, marker and trigger settings"
on page 892
Manual operation: See "Marker Mode" on page 259
[:SOURce<hw>]:BB:DM:TRIGger:OUTPut<ch>:PULSe:DIVider <Divider>
Sets the divider for pulse marker mode (PULSe).
*)
If R&S SMW-B9 is installed, the minimum marker duration depends on the sample/
symbol rate.
See "Marker Minimum Duration" on page 236.
Parameters:
<Divider> integer
Range: 2 (R&S SMW-B10) / 32* (R&S SMW-B9) to 1024
*RST: 2
Example: See Example "Adjusting clock, marker and trigger settings"
on page 892
Manual operation: See "Marker Mode" on page 259
[:SOURce<hw>]:BB:DM:TRIGger:OUTPut<ch>:PULSe:FREQuency?
Queries the pulse frequency of the pulsed marker signal PULSe.
Return values:
<Frequency> float
Example: See Example "Adjusting clock, marker and trigger settings"
on page 892
Usage: Query only
Manual operation: See "Marker Mode" on page 259