Remote Control Commands
R&S
®
SMW200A
1138User Manual 1175.6632.02 ─ 16
Value ranges
●
Queries return the current value of the respective register, which permits a check of
the device status.
Return values: A decimal value in the range 0 to 32767 (=2
15
-1)
●
The configuration commands set the respective register thus determining which
status changes of the R&S SMW cause the status registers to be changed.
Setting values: A decimal value in the range 0 to 32767 (=2
15
-1)
:STATus:OPERation:CONDition.................................................................................... 1138
:STATus:OPERation:ENABle........................................................................................ 1138
:STATus:OPERation[:EVENt]........................................................................................ 1138
:STATus:OPERation:NTRansition..................................................................................1139
:STATus:OPERation:PTRansition..................................................................................1139
:STATus:PRESet......................................................................................................... 1139
:STATus:QUEStionable:CONDition................................................................................1140
:STATus:QUEStionable:ENABle....................................................................................1140
:STATus:QUEStionable[:EVENt]....................................................................................1140
:STATus:QUEStionable:NTRansition............................................................................. 1140
:STATus:QUEStionable:PTRansition..............................................................................1141
:STATus:QUEue[:NEXT]?.............................................................................................1141
:STATus:OPERation:CONDition <Condition>
Sets the content of the CONDition part of the STATus:OPERation register. This part
contains information on the action currently being performed in the instrument. The
content is not deleted after being read out because it indicates the current hardware
status.
Parameters:
<Condition> string
Example:
:STATus:OPERation:CONDition?
queries the Status:Operation:Condition register.
:STATus:OPERation:ENABle <Enable>
Sets the bits of the ENABle part of the STATus:OPERation register. This setting deter-
mines which events of the Status-Event part are forwarded to the sum bit in the status
byte. These events can be used for a service request.
Parameters:
<Enable> string
Example:
:STAT:OPER:ENAB 32767
all events are forwarded to the sum bit of the status byte.
:STATus:OPERation[:EVENt] <Event>
Queries the content of the EVENt part of the STATus:OPERation register. This part
contains information on the actions performed in the instrument since the last readout.
The content of the EVENt part is deleted after being read out.
STATus Subsystem