Remote Control Commands
R&S
®
SMW200A
876User Manual 1175.6632.02 ─ 16
Example: See Example "Creating a frequency modulated RF signal"
on page 874.
Manual operation: See "Mode" on page 508
[:SOURce<hw>]:FM:RATio <Ratio>
Sets the deviation ratio (path2 to path1) in percent.
Parameters:
<Ratio> float
Range: 0 to 100
Increment: 0.01
*RST: 100
Example: See Example "Creating a frequency modulated RF signal"
on page 874.
Manual operation: See "Ratio Path2/Path1" on page 508
[:SOURce<hw>]:FM:SENSitivity?
Queries the sensitivity of the externally supplied signal for frequency modulation. The
sensitivity depends on the set modulation deviation.
Return values:
<Sensitivity> float
Sensitivity in Hz/V.
It is assigned to the voltage value for full modulation of the input.
Range: 0 to max
Increment: 0.01
Example: See Example "Using an external signal source" on page 874.
Usage: Query only
[:SOURce<hw>]:FM<ch>:SOURce <Source>
Selects the modulation source for frequency modulation.
Suffix:
FM<ch>
.
1|2
Modulation signal channel.
SOURce Subsystem

Table of Contents