Remote Control Commands
R&S
®
SMW200A
930User Manual 1175.6632.02 ─ 16
Example: See Example "Creating test signals" on page 920
Manual operation: See "Samples per Period" on page 313
[:SOURce<hw>]:BB:ARBitrary:TSIGnal:SINE:FREQuency <Frequency>
Sets the frequency of the simple sinusoidal test signal.
Parameters:
<Frequency> float
Range: 100 to depends on the installed options
Increment: 0.01
*RST: 1000
Default unit: Hz
Example: See Example "Creating test signals" on page 920.
Manual operation: See "Frequency" on page 311
[:SOURce<hw>]:BB:ARBitrary:TSIGnal:SINE:PHASe
<Phase>
Sets the phase offset of the sine wave on the Q channel relative to the sine wave on
the I channel.
Parameters:
<Phase> float
Range: -180 to 180
Increment: 0.01
*RST: 90
Default unit: DEG
Example: see Example "Creating test signals" on page 920
Manual operation: See "Phase Offset Q" on page 312
[:SOURce<hw>]:BB:ARBitrary:TSIGnal:SINE:SAMPles <Samples>
Sets the sample rate for the sine signal in samples per period.
The resulting clock rate must not exceed the maximum ARB clock rate (see data
sheet).
The maximum value is automatically restricted by reference to the set frequency and
has to fulfill the rule Frequency * Samples <= ARB clock rate.
Parameters:
<Samples> integer
Range: 4 to 1000
*RST: 100
Example: see Example "Creating test signals" on page 920
Manual operation: See "Samples per Period" on page 312
SOURce Subsystem