Remote Control Commands
R&S
®
SMW200A
976User Manual 1175.6632.02 ─ 16
Figure 13-9: Principle of creating a waveform manually
The following steps outline how to create the waveform file SICO.wv:
1. Calculate the sine and cosine values, e.g. use the SICO.cpp program.
The result is stored in the file SICO.txt.
Figure 13-10: Contents of SICO.txt: first column Sine (I), second column Cosine (Q)
2. Convert the values from the file SICO.txt into binary format consisting of 16-bit
signed integer numbers. The numeric range between –1.0 and +1.0 corresponds to
the modulation range of the waveform 16-bit D/A converter of -32767 to +32767.
+1.0 -> 32767 -> = 0x7FFF
0.0 -> 0 -> = 0x0000
-1.0 -> -32767 -> = 0x8001
SOURce Subsystem