Remote Control Commands
R&S
®
SMW200A
993User Manual 1175.6632.02 ─ 16
<Count> integer
number of carriers in the carrier range, starting from the
<Start> carrier
Range: 1 to max
Return values:
<Phas[,Phas..]> float
Example: see "Programming Examples" on page 983
Manual operation: See "Carrier Table" on page 378
[:SOURce<hw>]:BB:MCCW:CARRier:POWer
<CarrierIndex>, <Power>
Sets the power of the selected carrier.
Parameters:
<CarrierIndex> integer
Range: 0 to lastCarrier
<Power> float
Range: -80 to 0
Increment: 0.01
*RST: 0
Example: see "Programming Examples" on page 983
Manual operation: See "Carrier Table" on page 378
[:SOURce<hw>]:BB:MCCW:CARRier:LIST:POWer <Pow0[,Pow1..]>]
[:SOURce<hw>]:BB:MCCW:CARRier:LIST:POWer? <Start>, <Count>
Sets the power of the carrier with the aid of a value list.
Setting parameters:
<Pow0[,Pow1..]> float
Increment: 0.01
*RST: 0 dB
Query parameters:
<Start> integer
start carrier index
Range: 0 to lastCarrier
<Count> integer
number of carriers in the carrier range, starting from the
<Start> carrier
Range: 1 to lastCarrier
Return values:
<Pow[,Pow..]> float
Example: see "Programming Examples" on page 983